Technical Specification
Microprocessor Based with Printer Interface, Five elements measurement in single aspiration of the sample and Calibration Curve programming using maximum of 25 standards. |
Data processing through curve fitting techniques – least squares, quadratic, Multiple Calibration curves saving facility, Samples data storage facility, Printer attachment facility, Measurement in ppm & meq, useful for both medical & soil testing applications and setup storage facility Upto 10 setups. |
General Range: K : 0 – 100 ppm (Upto 250 meq/l, 1:100 dil.), Na : 0 – 100 ppm (Upto 250 meq/l, 1:100 dil.), Ca : 0 – 100 ppm (Upto 250 meq/l, 1:100 dil.), Li : 0 – 100 ppm (Upto 250 meq/l, 1:100 dil.) and Ba : 50 – 1000 ppm. Serum range (meq/l): Na: 0 – 200 meq/l, 1:100 dil., K: 0 – 10 meq/l, 1:100 dil. And Li: 0 – 2 meq/l, 1:5 dil. Urine Range (meq/l): Na: 0 – 250 meq/l, 1:100 dil., K: 0 – 100 meq/l, 1:100 dil. And Ca: 0 – 10 meq/l, 1:02 dil. |
Sensitivity: Na: 0.5 ppm, K: 0.5 ppm, Li: 0.5 ppm, Ca: 15 ppm and Ba: 50 ppm. |
Resolution: 0.1 ppm / 0.1 meq, Reproducibility: < 1% coefficient of variation for 20 consecutive samples using 10 ppm, Na set as maximum standard, Linearity: < 2% error when 3 ppm Na / K and 5 ppm Li are set as maximum standard, Curve fit Accuracy: ± 2 % of FS. |
Display: 20 x 4 character alphanumeric LCD display with backlit. |
Average time: In – Built in software. |
Flame system: LPG & Dry oil free air, Detector: Photodiode. |
Gas control: Adjustable with knob. |
Ignition system: Auto Ignition |
Automiser: Axial Flow Type. |
Auto Gas Cutoff facility (Model BGS – 316 only), Auto Flame Failure Detection Facility (Model BGS – 316 only). |
Optional Accessories: PC software at extra cost and Ca & Li are optional at extra cost. |